Software User guide

Settings 9
Con gXP Users Manual
Using the User Units Wizard
The User Units Wizard operates in two modes: Linear Fit or Tank Level. Both modes calculate an optimum User Factor and User O set.
The Linear Fit tab
The Linear Fit tab allows you to quickly calculate the best User Factor and User O set by entering actual measurements and the corresponding desired read-
ing for each measurement. In each row, enter the displayed pressure in the  rst column and the reading you want the gauge to display in the Desired Indica-
tion column. The User Factor and User O
set will be recalculated after each point entered. The Gauge Reading column shows what the gauge will actually
display using the calculated User Factor and User O set. The Delta column indicates how much error there is between the Desired Indication and the Gauge
Reading
.