Safety & Certications 33
XP2i Operation Manual
Safety Instructions for Hazardous Locations – ENGLISH (English)
•
Do not use the RS-232 connector in a hazardous location.
•
Replace batteries in non-hazardous locations, with approved batteries, only.
•
It is the users responsibility to understand the proper application of this product in potentially explosive atmospheres.
X
Approved Batteries – ENGLISH (English)
The XP2i is Intrinsically Safe only if powered by one of the following battery types:
Approved Battery Type Ta= Marking
Rayovac Max Plus 815
-20 to 50° C Ex ia IIC T4 Ga
Energizer E91
Energizer EN91
-20 to 50° C Ex ia IIC T3 Ga
Duracell MN1500
Replace batteries with approved type in non-hazardous locations only
Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety
–
do not assume other models are equivalent.
Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.
Instrucciones De Seguridad Para Zonas Peligrosas – ESPAÑOL (Spanish)
•
No use el conector RS-232 en zona clasicada.
•
Cambie las pilas en zona no clasicada, solo con pilas aprobadas.
•
Es responsabilidad del usario comprender la aplicación de este producto en atmósferas potencialmente explosivas
X
Pilas Aprobadas – ESPAÑOL (Spanish)
El XP2i solo es intrínsecamente seguro si se alimenta con uno de los siguientes tipos de pilas:
Approved Battery Type Ta= Marking
Rayovac Max Plus 815
-20 to 50° C Ex ia IIC T4 Ga
Energizer E91
Energizer EN91
-20 to 50° C Ex ia IIC T3 Ga
Duracell MN1500
Replace batteries with approved type in non-hazardous locations only
Se han probado muchos otros tipos de baterías pero han fallado el cumplimiento de los requisitos para la seguridad intrínseca
–
No asuma que otros modelos
son equivalentes.
Energizer está fabricado por Energizer Holdings, Inc., y por Eveready Battery Company, Inc.