Owner manual

PN: 1763 Rev C, 4/2004
Page 6 of 8
PRESSURE is Our BUSINESS
Pressure
Range Accuracy
0 to 200 kPa ±(0.05% of Reading + 0.1H
2
O)
0 to 30 PSI ±(0.05% of Reading + 0.01PSI)
30 to 50 PSI ±(0.25% of Reading)
0 to -100kPa ±(0.5% of Reading + 0.14 kPa)
0 to -14.7 PSI ±(0.25% of Reading), typical
Maximum Working Pressure: 100 PSI
Burst Pressure: 500 PSI
Wetted Materials: 316 Stainless Steel, PTFE impregnated hard
anodized aluminum, and buna-n rubber.
Unless otherwise specied at time of purchase: 1 PSI = 27.6806H
2
O (water at 4°C)
Operating Environment
Temperature Range Humidity
-10 to 10°C Uncontrolled Humidity
10 to 30°C 0 to 95% Relative Humidity
30 to 40°C 0 to 75% Relative Humidity
40 to 50°C 0 to 45% Relative Humidity
50 to 55°C 0 to 35% Relative Humidity
Temperature Derating
Add to Basic Accuracy Specification. °C = ambient temperature.
Temperature Range Derating
28 to 55°C (0.016%/°C) x (°C - 28°C)
18 to 28°C No Derating
0 to 18°C (0.048%/°C) x (18° - °C)
-10 to 0°C (0.264%/°C) x (9° - °C)
General Specifications
Weight 164 grams (5.8 oz.) w/battery
Overall length 118mm (4
5
/
8
)
Battery NEDA#1604, 6F22, 006P
Battery Life 400 Hours typ. w/alkaline battery
Output 1mV/unit into 10MΩ load
Storage -51°C to 71°C