nVision Operation Manual for Reference Recorder & Lab Reference
Contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Quickstart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview 1 —— Overview INTRODUCTION Thank you for choosing the nVision Reference Recorder from Crystal Engineering Corporation. The philosophy behind nVision: nVision lets you visualize measurements graphically, with or without a pc, in real time as it is being recorded. It is much easier to identify trends or anomalies visually, than in tables of data or spreadsheets. nVision is tremendously flexible and can be configured to measure and record a variety of combinations of measurements.
—— Overview 2 QUICKSTART nVision Reference Recorder shown.
—— Functions 3 Functions ON/OFF Power button Press and hold the ( power) button for 1 second to turn the nVision on or off. The nVision will automatically power down if not used for the time period defined in CrystalControl. Automatic Shutoff - Low Power Mode Adjust your Automatic Shutoff time (shut off time in absence of key press) to optimize battery life. This feature is adjustable from 30 seconds to “always on.” During a recording, the nVision will enter Low Power Mode instead of shutting off.
Functions 4 Zero To zero the nVision: • Press the (zero) button for at least 1 second while vented to atmosphere until the dashed lines (- - - - -) appear. To clear the zero value: • Hold the (zero) button for 5 seconds until the display changes from (- - - - -) to the zeroed value, then to (- - -). You can adjust the Zero Limit at which the (zero) button will display “- -HI- -” in CrystalControl.
—— Functions 5 NUMERICAL DISPLAY OVERVIEW Numerical Upper and Lower Module screens High / Low: Peak always measured at 10 readings per second Upper Module Screen Lower Module Screen Vertical Scale: Applied versus full-scale pressure (Pressure Module Only) —— Data Point Counter Screen Dual Module Screen Data Point Screen (Single Sensor) Note: To change the displayed units on any Numerical Display, see the Units section.
—— Functions 6 GRAPHING Navigating the Graphical Display In the graphical modes the nVision navipad enables you to control how you view your data. The ( ) & () keys allow you to navigate to specific points along your run, while displaying reading and time information. The () & () keys allow you to zoom in and out of your recorded run to suit your needs.
—— —— Functions 7 AVERAGING MODE Averaging mode reports the average reading during the recorded run. If this screen is enabled, data displayed here represents the average of all past data points, over the duration of a recording. The start date and time, duration of the recorded run, and the live reading are displayed. nVision calculates averages as follows: r1 + r2 + r3 ...
Functions 8 Tare Using the Tare function improves your differential measurement uncertainty significantly if used properly. The Tare function equalizes (normalizes) the nVision’s two modules at a non-ambient datum. The Tare reading displays the same units as the main ΔP or ΔT reading. If you apply the same static line pressure, temperature, or resistance signal to both sensors simultaneously, you should have a differential reading of zero.
Chassis 9 —— Chassis CHASSIS CONTROLS Setup Button Pressing the (setup) button brings up a selectable menu including Clear Peaks, Units, Settings, Summary, and Recording. On the (navipad), use the () and () arrows to move to the desired feature and use the (select) button or () and () arrows to move into the desired function. The nVision Setup menu Clear Peaks (Resetting Hi and Lo Peaks) Note: Dashed lines will briefly appear across the peak value indicators.
Chassis 10 XX Changing Units—Single Module, Numerical or Graphical Screen XX Changing Units—Dual Module, Numerical or Graphical Screen XX Changing Units—Differential Screen XX Changing Units—BARO, Numerical or Graphical Screen XX Changing Between Absolute and Gauge Pressure—Numerical or Graphical Screen nVision Operation Manual
Chassis 11 Settings XX Adjust Contrast (100%) (75%) (50%) XX Lock the nVision The Screen Lock Password feature will also be found in Settings, if enabled in CrystalControl. Prevent access to your nVision by protecting your device with a 4-digit keypad lockout set in CrystalControl. To unlock the nVision, simply enter the 4 digit password with the arrow keys and press the (select) button. The correct password will allow you back into standard nVision operation.
Chassis 12 Summary Screen The Summary screen allows you to view details or settings in your chassis or module. XX View a Module Summary Once you have selected a summary screen, you may also use the () and ( ) arrows to move to the next summary screen. Press the (Setup) button again to exit the summary screen. Note: If a BARO module is installed, BARO Module will appear in the Summary drop-down menu list below Lower Module.
Chassis 13 Recording XX Start a Recording Press the (record) button until the red LED indicator flashes. XX View the current Run Tag during a recording XX Erase All Runs A confirmation screen will ask, Are You Sure?. Press (select) to continue and erase all the data runs on this nVision. Press the (back) button to cancel. —— !!CAUTION: Never remove power (either battery or USB) during the erasing process.
—— Chassis 14 NVISION REFERENCE RECORDER SPECIFICATIONS Temperature (Operating and Storage) ALL DIMENSIONS ARE IN MILLIMETERS [INCHES] Operating & Compensated ������������� -20 to 50°C (-4 to 122°F). RTD100 MODULE SHOWN WITH INCLUDED PROTECTIVE CAP Storage ��������������������������������������������������� - 40 to 75°C (-40 to 167°F). RTD100 MATING CONNECTOR FOR RTD PROBE SHOWN INSTALLED 30 1.20 2 PLACES 15 0.60 RTD100 CONNECTOR SHOWN WITHOUT PROTECTIVE CAP 51 2.00 110 [4.
—— Chassis 15 NVISION LAB REFERENCE SPECIFICATIONS Temperature (Operating and Storage) 294 [11.6] Operating & Compensated ������������� 10 to 50°C (50 to 122°F). 150 [5.9] MAXIMUM WITH TWO PRESSURE MODULES 280 [11.02] Storage ��������������������������������������������������� 0 to 75°C (32 to 167°F). 256 [10.1] 143 [5.6] UPPER PORT PRESSURE MODULE (PM) SHOWN CRYSTAL CPF SYSTEM: WITH 7/16"-20 INTERNAL THREADS) Humidity 120 [4.
—— Modules 16 Modules MODULE INSTALLATION INSTRUCTIONS The nVision’s upper and lower bays allow for removal of modules in the field. All module changes should be completed in a dry, clean environment, indoors. Proper electrostatic discharge (ESD) grounding techniques should be taken into account prior to the module change over. If you’re removing a module without installing a replacement, a blank plate (P/N: BNKPLT) must be installed to ensure your IP67 rating and to protect the product.
Modules 17 5 Install a new module in the same orientation as the one removed in step 4. The tri-lobe design of the module will not allow improper installation; do not force installation of the module as permanent damage may occur. To ensure an IP67 rated seal, lightly lubricate the module’s o-ring (P/N: 4110) with Dow 111 silicon lubricant or equivalent. (4x) Lightly lubricate the o-ring, then install the module in the proper orientation.
—— Modules 18 PRESSURE MODULE (PM) INSTRUCTIONS Pressure Connection Crystal CPF System: Medium Pressure Female (MPF) (1/4” medium pressure tube system with 7/16-20 threads). See our CPF Brochure for further information. CPF o-ring size and material: AS568A-012, Viton 90 durometer (P/N 3981). For most applications CPF Fittings can be hand tightened (no tools required).
—— Modules 19 PRESSURE MODULE (PM) SPECIFICATIONS Accuracy (Gauge) Pressure Module Tables PSI Module Range 0 - 30% Gauge Full Scale 30 - 110% Gauge Full Scale PSIG ± (% Full Scale) ± (% of Reading) 30 0.0075% 0.025% 100 0.0075% 0.025% 300 0.0075% 0.025% 1000 0.015% 0.05% 3000 0.015% 0.05% 10000 0.03% 0.1% 15000 0.03% 0.1% barG ± (% Full Scale) ± (% of Reading) 0 to 30% of Full Scale ������������������������������ ±(0.0075% of Full Scale), ±(0.
Modules 20 Accuracies, Ranges, and Resolutions PSI bar kPa/MPa kg/cm2 30PSI 0.025% of Reading modules 3BAR 300KPA 3KG 100PSI 10BAR 1MPA 10KG 300PSI 30BAR 3MPA 30KG 1KPSI 0.05% of Reading modules 100BAR 10MPA 100KG 3KPSI 300BAR 30MPA 300KG 10KPSI 0.1% of Reading modules 700BAR 70MPA 700KG 15KPSI 1000BAR 100MPA 1000KG Overpressure PSI kg/cm2 inHg inH20 * mmHg mmH20 kPa bar mbar 3.0 x 3.0 x 3.0 x 3.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 2.0 x 1.5 x 1.5 x 1.
Modules 21 Differential Pressure Measurement Uncertainties without Tare The total nVision Reference Calibrator measurement uncertainty in the ΔP mode configuration will need to consider the uncertainties of both pressure modules. We recommend the module uncertainties to be combined with the preferred square root of the sum of the squares (or “root sum squares”) method. The following table lists the possible combinations of combining Pressure Modules (PM) with different accuracy statements.
Modules 22 Differential Pressure Resolution PSI bar kPa/MPa kg/cm2 PSI kg/cm2 inHg inH2O mmHg mmH2O kPa bar mbar MPa 30 3 300 3 0.0001 0.00001 0.0001 0.001 0.001 0.1 0.001 0.00001 0.01 0.00001 100 10 1 10 0.0001 0.00001 0.001 0.01 0.01 0.1 0.001 0.00001 0.01 0.00001 300 30 3 30 0.001 0.0001 0.001 0.01 0.01 0.1 0.01 0.0001 0.1 0.00001 1000 100 10 100 0.01 0.0001 0.01 0.1 0.1 0.1 0.01 0.0001 0.1 0.00001 3000 300 30 300 0.01 0.
—— Modules 23 BAROMETRIC REFERENCE (BARO) MODULE INSTRUCTIONS Installation of the BARO module in the power bay allows you to convert your gauge pressure measurement to the absolute scale if so desired. You may also view the barometric reference reading directly on a dedicated numerical screen. When the BARO module is installed, it may also be disabled in CrystalControl to conserve battery power, however your nVision will not be able to display absolute readings unless it is active.
—— Modules 24 BAROMETRIC REFERENCE (BARO) MODULE SPECIFICATIONS Absolute pressure mode is achieved by using the BARO Reference Module to establish a datum. The PM and BARO sensor uncertainties were combined to establish a new accuracy statement for readings taken in the absolute mode. It must be noted that the accuracy statement is valid for readings of 1 barA or greater. Accuracy ± 0.00725 PSI, ± 0.
—— Modules 25 ABSOLUTE PRESSURE SPECIFICATIONS PSIA (Pressure with BARO module) barA (Pressure with BARO module) MPaA (Pressure with BARO module) 30 PSI module 3 bar module 300 kPa module 0.200 to 14.500 PSIA: ±0.011 PSIA, typical 14.500 to 44.500 PSIA: ±(0.025% of Reading) + 0.003 PSIA 100 PSI module 0.200 to 14.500 PSIA: ±0.011 PSIA, typical 14.500 to 44.500 PSIA: ± 0.011 PSIA 44.500 to 114.500 PSIA: ±(0.025% of Reading) 300 PSI module 0.20 to 14.50 PSIA: ±0.01 PSIA, typical 14.50 to 104.
—— Modules 26 CURRENT, VOLTAGE, AND SWITCH TEST (MA20) MODULE INSTRUCTIONS The nVision MA20 module has three operational modes: current measurement, voltage measurement, and switch test. Each mode may be selected via the Setup menu, and can only be operated one at a time. To ensure proper connection to the MA20 Module use the following strategy: 1 Ensure that power is off on the circuit that you are about to measure.
Modules 27 XX Current Measurement The nVision may be used to measure current up to 50mA. Select the desired current mode through the Setup menu to properly configure the nVision prior to connection and use. XX Current Measurement with HART Resistor The nVision may be used to measure current in a circuit that includes a HART transmitter or device. For devices that use the HART protocol, a load resistor must be placed in the loop. The HART input on the MA20 provides a 250 Ohm load resistor.
Modules 28 Switch Test Mode The nVision may be used to detect switch closures. Select the Switch Test mode through the Setup menu to properly configure the nVision prior to connection and use. The illustration denotes the proper Switch Test connection scheme to the MA20 module. —— MA20 module Switch Test connection.
Modules 29 Logging Interval Fastest Logging Interval. . . . . . . 6 readings per second. Note: Although nVision logging interval may be set to a faster rate, the MA20 module will update the reading at 6 times per second. ATEX and IECEx Scheme Entity Parameters The MA20 Module has these specific input entity parameters: Ui = 28 V Uo = 6.6 V li = 93.3 mA lo = 4.45 mA Pi = 653.3 mW Po = 7.34 mW Ci = 0.36 uF Co = 0.5 uF** Li = 39.
Modules 30 Voltage (V) Input XX Accuracy ±(0.015% of reading + 0.002VDC) Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range for one year. Note: Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.
Modules 31 Connecting your RTD to the RTD100 Module Your nVision RTD100 module has been shipped with an IP67 rated, M8 connector (P/N: 3953). The terminal block based connector allows you to attach your RTD sensor for 2-, 3-, or 4-wire connections. Care must be taken to install the RTD connector shell properly to ensure the robust IP67 sealing. 1 RTD sensor element connection requirements: • 100Ω Platinum RTD with: • • • TCR of 385, 3926, or 3911 Cable diameter of 3.5 to 5.0mm [0.14 to 0.
Modules 32 (d) Install your RTD sense element wires as appropriate for your configuration (see illustration below), and tighten the set screws. (e) Confirm correct orientation of element wires in connected state. 2-wire 3-wire 4-wire 4 Lubricate o-ring and thread system with Dow 111 or equivalent in location near terminal block as required to prevent water intrusion when connector shell is installed. Thread shell in place until tight. 5 Leak check to ensure water tight seal.
Modules 33 To calculate the total uncertainty of the differential temperature measurement (UΔT ) you must combine the uncertainties of the upper and lower RTD100 and RTD sensor element systems (Usystem).
Modules 34 Logging Interval Fastest Logging Interval ������������������� 5 readings per second. Note: Although the nVision logging interval may be set to a faster rate, the RTD100 module will update the reading at 5 times per second. ATEX and IECEx Scheme Entity Parameters The RTD100 module has these specific input entity parameters: Ui = 0 Uo = 9.73 V li = 0 lo = 1.6642 A Pi = 0 Po = 1.1 W Co = 0.
Modules 35 RTD100 System Measurement Uncertainties To understand the total system measurement uncertainty of the temperature measurement you must consider both the nVision and the RTD sensing element uncertainties utilized in the test application. Since the uncertainties of nVision and the sense element are independent of each other, they must be combined properly with the preferred square root of the sum of the squares1 (or “root sum squares”) method.
—— Power 36 Power BATTERY POWER (REFERENCE RECORDER ONLY) Power Icon States The nVision Reference screen display has the following power icon states: Power Icon Key - nVision Reference Recorder Icon Description External Power (USB) 100% 75% 50% 25% 0%* Power Remaining * Replace Batteries or connect to USB Power The icon will appear when the batteries are exhausted and will need to be changed to ensure full functionality of the nVision.
Power 37 Extending Battery Life and Low Power Modes Note that the nVision reference recorder has many customizable battery saving features available to you for optimization in CrystalControl. XX Low Power Mode You may extend battery life substantially by slowing down your Logging Interval (recording rate), reducing the Automatic Shutoff and Backlight Shutoff times, or reducing the number of screens enabled.
Power 38 XX The nVision is CSA certified only if powered by one of the following battery types: Approved Battery Type Ta= Rayovac Max Plus 815 -20 to 50° C Duracell MN1500 -20 to 45° C Energizer E91 Energizer EN91 Duracell MN1500 Marking Class I, Division 1, Grp C, D T4 Class I, Division 1, Grp C, D T3B -20 to 50° C Class I, Division 1, Grp C, D T3A Class I, Division 1, Grp C, D T3C Replace batteries with approved type in non-hazardous locations only XX Battery Life Settings such as Auto Shut
Safety and Certifications 39 —— Safety and Certifications HAZARDOUS LOCATIONS Note: The approvals apply only to the nVision Reference Recorder. The nVision reference recorder includes the following Intrinsic Safety approvals: II 1G Ex ia IIB T4 Ga -or- Ex ia IIB T3 Ga Intrinsically Safe and Non-incendive for Hazardous Locations: Ta = -20C to 50C or -20C to 45C (depending on type of approved battery used) CSA Class I, Division 1, Groups C and D, Temperature Code T4 / T3A / T3B / T3C.
Safety and Certifications 40 ATEX EC Declaration of Conformity PED EC Declaration of Conformity Note: These approvals apply only to the nVision Reference Recorder. Note: These approvals apply to both the nVision Reference Recorder & Lab Reference. DECLARATION OF CONFORMITY DECLARATION OF CONFORMITY In accordance with the ATEX Directive 94/9/EC According to ISO/IEC 17050-1:2004 Manufacturer’s Name: Crystal Engineering Corporation An AMETEK Inc.
Safety and Certifications 41 EMC EC Declaration of Conformity EMC EC Declaration of Conformity Note: These approvals apply only to the nVision Reference Recorder. Note: These approvals apply only to the nVision Lab Reference. DECLARATION OF CONFORMITY According to ISO/IEC 17050-1:2010 Manufacturer’s Name: Crystal Engineering Corporation An AMETEK Inc.
—— Safety and Certifications 42 MULTI-LANGUAGE SAFETY INSTRUCTIONS Česky (Czech) XX Bezpečnostní instrukce pro prostředí s nebezpečím výbuchu • V prostředí s nebezpečím výbuchu nepoužívejte přípojku USB. • Baterie vyměňujte pouze v bezpečném prostředí. Používejte pouze schválené baterie. • Za správné použití tohoto přístroje v prostředí s nebezpečím výbuchu odpovídá jeho uživatel.
Safety and Certifications 43 English (English) XX Safety Instructions for Hazardous Locations • Do not use the USB connector in a hazardous location. • Replace batteries in non-hazardous locations, with approved batteries, only. • It is the users responsibility to understand the proper application of this product in potentially explosive atmospheres.
Safety and Certifications 44 Français (French) XX Instructions de sécurité pour les Zones Dangereuses • Ne pas utilisez le connecteur USB dans une Zone Dangereuse. • Remplacez les piles dans des Zones non-dangereuses, avec les piles appropriées, uniquement. • Il est de la responsabilité de l’utilisateur de bien comprendre l’application appropriée de ce produit en atmosphères explosives.
Safety and Certifications 45 Nederlands (Dutch) XX Veiligheidsinstructie voor gebruik in een explosie gevaarlijkeomgeving • Het gebruik van de USB interface is niet toegestaan in een explosie gevaarlijke omgeving. • Vervang de batterijen uitsluitend in een niet explosie gevaarlijke omgeving en gebruik alleen batterijen welke zijn goedgekeurd en toegestaan. • De gebruiker dient er mee bekend te zijn welke gevaren er kunnen optreden in een explosie gevaarlijke ruimte bij gebruik van dit product.
Safety and Certifications 46 Suomen Kieli (Finnish) XX Räjähdysvaarallisten tilojen turvallisuusohjeita • USB väylää/liitintä EI saa käyttää räjähdysvaarallisissa tiloissa. • USB väylää/liitintä EI saa käyttää räjähdysvaarallisissa tiloissa. Käytettävä ehdottomasti ja ainoastaan hyväksyttyjä paristoja. • Käyttäjän vastuulla on laitteen käyttö räjähdysvaarallisissa tiloissa. Mittausovellus ja käyttöympäristö on ehdottomasti selvitettävä ennen käyttöä.
—— Support 47 Support TROUBLESHOOTING The nVision is a very high performance reference recorder. Due to the high resolution of this product, you may observe conditions that appear to be defects in the product, but are in fact a result of being able to read and measure pressure to a degree not possible with other instruments.
Support 48 Error 1 displayed XX Reason The nVision checks the integrity of internal calibration coefficients every time it’s turned on. If any coefficients have been corrupted in any way, “Error 1” is displayed. XX Solution Contact factory for instructions on how to restore the memory to the original factory settings. Error 2 displayed XX Reason The nVision has tried to display a number too large for the display (i.e., more than 6 digits). May be due to an electrical malfunction or numerical error.
—— Support 49 CALIBRATION If adjustment is required, we recommend returning the nVision or separate modules to the factory. Factory service offers benefits you won’t find anywhere else. Factory calibration tests your nVision at a variety of temperatures utilizing NIST traceable standards, resulting in calibration certificates that provide performance data over temperature. Our calibration facilities are A2LA accredited to ISO 17025:2005 & ANSI/NCSL Z540-1-1994.
Support 50 Calibration for Current, Voltage & Switch Test Module (MA20) The MA20 can be calibrated through the use of the span factor (userspan) and the Offset for the current and voltage modes. Record displayed current or voltage for two or more points. Determine if the nVision would benefit from an overal increase or decrease of the indicated reading and modify using the userspan and offset feature of CrystalControl.
—— —— —— Support 51 CONTACT US Phone ��������������������������������������������������� (805) 595-5477 Toll-Free ����������������������������������������������� (800) 444‑1850 Fax ����������������������������������������������������������� (805) 595-5466 Email ����������������������������������������������������� crystal@ametek.com Web ��������������������������������������������������������� crystalengineering.
© 2014 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 4701.