User Manual
Modules 19
nVision Operation Manual
—
PRESSURE MODULE PM SPECIFICATIONS
Pressure Module Tables
Module
Range
0 - 30%
Gauge Full Scale
30 - 110%
Gauge Full Scale
PSIG ± (% Full Scale) ± (% of Reading)
PSI
30 0.0075% 0.025%
100 0.0075% 0.025%
300 0.0075% 0.025%
1000 0.015% 0.05%
3000 0.015% 0.05%
10000 0.03% 0.1%
15000 0.03% 0.1%
barG ± (% Full Scale) ± (% of Reading)
bar
3 0.0075% 0.025%
10 0.0075% 0.025%
30 0.0075% 0.025%
100 0.015% 0.05%
300 0.015% 0.05%
700 0.03% 0.1%
1000 0.03% 0.1%
kPaG / MPaG ± (% Full Scale) ± (% of Reading)
kPa /
MPa
300 0.0075% 0.025%
1 0.0075% 0.025%
3 0.0075% 0.025%
10 0.015% 0.05%
30 0.015% 0.05%
70 0.03% 0.1%
100 0.03% 0.1%
kg/cm
2
G ± (% Full Scale) ± (% of Reading)
kg/cm
2
3 0.0075% 0.025%
10 0.0075% 0.025%
30 0.0075% 0.025%
100 0.015% 0.05%
300 0.015% 0.05%
700 0.03% 0.1%
1000 0.03% 0.1%
Accuracy (Gauge)
0 to 30% of Full Scale . . . . . . . . . . . . . . . ±(0.0075% of Full Scale), ±(0.015% of Full Scale), or ±(0.03% of Full Scale)
30 to 110% of Full Scale . . . . . . . . . . . . . ±(0.025% of Reading), ±(0.05% of Reading), or ±(0.1% of Reading)
Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . . . . for 300PSI, 30BAR, 3MPA, 30KG and lower pressure modules:
±(0.06% of Full Scale*, typical)
*Full Scale = -14
.5 PSIG, -1.0 bar, -99.9 kPa, -1.0 kg/cm
2
.
Accuracy specications include all eects of linearity, hysteresis, repeatability, temperature, and stability for one year.
Note: Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent
recertication period.
P
M modules must be exercised and re-zeroed whenever exposed to signicant changes in environmental conditions to achieve these
specications. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest.
A properly exercised gauge will return to a zero reading (or return to the same ambient barometric reading).
!
CAUTION: Pressure Modules (PM) are not recommended for continuous use at high vacuum.