GaugeCalHP Comparator Operation Manual
Support
TROUBLESHOOTING
Unstable Pressure
X
Problem: Pressure cannot be generated correctly, or set pressure does not remain stable.
X
Solutions:
• Check that all adapters and pressure ttings have been tightened suciently to eliminate leaks.
• Conrm test uid does not contain air (air in the uid can increase adiabatic eects). You may have to adjust the pressure a few times before the
pressure is stable enough for calibration.
Replace test uid if necessary.
• Adiabatic and thermodynamic eects may be aecting the pressure reading. High pressure hydraulic systems are sensitive to temperature and move-
ments.
This eect will disappear after a period of time.
If the pressure is still not stabilizing, the system may be in need of service. Please contact your local distributor for advice.
Note: Always readjust the pressure after 5 to 10 minutes to allow time for settling of above eects.
MAINTENANCE
Users should/must carry out the following cleaning procedures when required:
Standard exterior cleaning . . . . . . .Clean using a soft, damp cloth.
Heavily soiled cleaning . . . . . . . . . . . The instrument may also be cleaned using isopropyl alcohol when heavily soiled.
Reservoir . . . . . . . . . . . . . . . . . . . . . . . . Can be removed for easy cleaning. Turn the GaugeCalHP onto its side and loosen the two screws holding the reservoir in
place. Slide the screws through the hole and quickly remove the reservoir as uid may leak out of the bottom.
To clean the clear acrylic reservoir tube use a soft cloth or soap and water only. Never use chemical cleaners.