User guide

Operating Instructions 6
P014 Hydraulic Jack Pump Operation Manual
Operating Instructions

GENERATING PRESSURE
Follow these steps to use the pump.
1 Lower the handle to its lowest position.
2 Loosen the adjustment screw and turn the banjo set to the desired position. Firmly tighten the screw.
3 Remove plugs using a 6mm Allen key, and ll with test uid
(
if necessary).
4 Connect the optional extension tube, pressure hose, and other connection ttings, as needed.
!
CAUTION: USE BONDED SEALS only, for attachment of any part or object directly to the manifold. Tighten all connections rmly.
5 Pump the handle gently until test uid is evident at the top of the connection ttings.
6 Connect the reference indicator and the device under test. Tighten the threads rmly.
Note: If CPF ttings are used in your pressure system, remaining air can be purged through the weep hole, by applying 10 psi or less and loosening the CPF
tting closest to the pressure device about two and a half turns. Repeat this process for each pressure device in the pressure system which is connected
to a CPF tting.
7 Power on the reference indicator and the device under test.
8 Conrm there is no pressure in the system by turning the vent valve.
9 Zero the reference indicator and the device under test.
10 Close the vent valve.
11 Verify the system connections are tightened rmly and pressurize the system by operating the pump handle gently until the approximate pressure
is obtained.
12 Allow time for the adiabatic eect to stabilize.
13 Use the ne adjust knob, if installed, to obtain the exact calibration pressure.
14 After obtaining your readings, release pressure by opening the vent valve.
!
WARNING: The device under test and the reference instrument should only be removed when there is no pressure in the system.
Note: AMETEK Denmark’s liability ceases if: Parts are replaced/repaired using spare parts which are not identical to those recommended by the
manufacturer or non-original parts are used in any way when operating the system. AMETEK Denmarks liability is restricted to errors, which
originated from the factory.