User guide
Operating Instructions 6
P014 Hydraulic Jack Pump Operation Manual
Operating Instructions
GENERATING PRESSURE
Follow these steps to use the pump.
1 Lower the handle to its lowest position.
2 Loosen the adjustment screw and turn the banjo set to the desired position. Firmly tighten the screw.
3 Remove plugs using a 6mm Allen key, and ll with test uid
(
if necessary).
4 Connect the optional extension tube, pressure hose, and other connection ttings, as needed.
!
CAUTION: USE BONDED SEALS only, for attachment of any part or object directly to the manifold. Tighten all connections rmly.
5 Pump the handle gently until test uid is evident at the top of the connection ttings.
6 Connect the reference indicator and the device under test. Tighten the threads rmly.
Note: If CPF ttings are used in your pressure system, remaining air can be purged through the weep hole, by applying 10 psi or less and loosening the CPF
tting closest to the pressure device about two and a half turns. Repeat this process for each pressure device in the pressure system which is connected
to a CPF tting.
7 Power on the reference indicator and the device under test.
8 Conrm there is no pressure in the system by turning the vent valve.
9 Zero the reference indicator and the device under test.
10 Close the vent valve.
11 Verify the system connections are tightened rmly and pressurize the system by operating the pump handle gently until the approximate pressure
is obtained.
12 Allow time for the adiabatic eect to stabilize.
13 Use the ne adjust knob, if installed, to obtain the exact calibration pressure.
14 After obtaining your readings, release pressure by opening the vent valve.
!
WARNING: The device under test and the reference instrument should only be removed when there is no pressure in the system.
Note: AMETEK Denmark’s liability ceases if: Parts are replaced/repaired using spare parts which are not identical to those recommended by the
manufacturer or non-original parts are used in any way when operating the system. AMETEK Denmark’s liability is restricted to errors, which
originated from the factory.