Operation Manual for XP2i-DP Digital Test Gauge
Contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Operating Instructions. . . . . . . . . . . . . . . . . . . . . . . . . .
—— —— Overview 1 Overview Introduction Thank you for choosing the XP2i-DP Digital Test Gauge from Crystal Engineering Corporation. The XP2i-DP is a combination of leading edge technology, double-sided isolated differential pressure sensing, and rugged industrial design. Based on the XP2i Series platform, the –DP has all the key features of the product line with additional capabilities that include: • • 0.
Specifications 2 Specifications Differential Pressure Accuracy 20 to 100% of Full Scale ������������������� ± (0.1% of Reading + Static Line Effect). 0 to 20% of Full Scale ����������������������� ± (0.02% of Full Scale + Static Line Effect). Static Line Effect: 0.0003 psi/psi of lowest static pressure. Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Temperature Operating ��������������������������������������������� -10 to 50° C (14 to 122° F).
Specifications 3 Connections Pressure Connections ����������������������� 1/8" Female NPT (both + and – ports). Communication ��������������������������������� DB9, RS-232 (environmentally sealed). Note: Do not use the RS-232 connector in a hazardous atmosphere. Display Screen ����������������������������������������������������� 5.5 digits. Display Rate ����������������������������������������� 4 readings/second (standard). 8 readings/second (PSV mode). Numerical Display Height ��������������� 16.
Specifications 4 Approved Batteries The XP2i-DP is Intrinsically Safe only if powered by one of the following battery types: Approved Battery Type Rayovac Max Plus 815 * Energizer EN91* Energizer E91 Duracell MN1500 Ta= Marking -20 to 50° C Ex ia IIB T4 Ga -20 to 50° C Ex ia IIB T3 Ga Replace batteries with approved type in non-hazardous locations only * Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.
—— Specifications 5 Pressure Ranges, Display Scales, and Resolution psi bar kPa MPa Overpressure (+/– ports) psi inH20 inHg mmHg mmH20 kg/cm2 bar mbar kPa MPa 15PSI 1BAR 100KPA 0.1MPA 6.5 x / 3.0 x 0.001 0.01 0.001 0.01 1 0.0001 0.0001 0.1 0.01 0.00001 100PSI 7BAR 700KPA 0.7MPA 2.0 x / 2.0 x 0.01 0.1 0.01 0.1 1 0.0001 0.0001 0.1 0.01 0.00001 • • • • Unneeded pressure units may be disabled via the RS-232 connector using ConfigXP software.
—— Applications 6 Applications Differential Pressure The XP2i-DP may be used to determine accurate differential pressure measurements across a range of applications or installations such as filters, orifice plates, valves, etc. For example, in a piping system with an upstream static pressure of approximately 80 psig, the XP2i-DP may be used to determine the 10 psid pressure drop across the device under test (DUT). In this case, the product would be connected as shown below.
—— Applications 7 Continuous Vacuum The unique design of the differential pressure sensor integrated into the XP2i-DP allows for long term vacuum use without fear of permanent damage to the sensing chip or package. When measuring high vacuum gauge pressures continuously it is recommended to connect the static line pressure to the right (–) port marked "Vacuum this side". To achieve gauge measurements, the left (+) port is left vented to atmospheric pressure.
Safety & Certifications 8 —— Safety & Certifications Hazardous Locations Every XP2i-DP pressure gauge includes the following Intrinsic Safety approvals: II 1G Ex ia IIC T4/T3 Ga ATEX Intrinsically Safe and Non-incendive for Hazardous Locations: Class I, Division 1, FTZU 12 ATEX 0048X CSA Groups A, B, C and D, Temperature Code T4 or T3C. This product conforms to the following standards: EN 60079-0 : 2009 | EN 60079-11: 2012 | EN 60079-26: 2007 Ex ia IIC T4/T3 Ga IECEx FTZU 12.
—— ——— — Support 9 Support Software ConfigXP Configuration Software Use ConfigXP to disable unwanted pressure units, set default pressure units, change water density, adjust calibration, and more via the RS-232 interface. DataLoggerXP Software DataLoggerXP is an optional upgrade to transform the XP2i-DP into a pressure datalogging device. Replacement Parts The only user-replaceable parts are the batteries.
—— Support 10 Accessories P/N 2984 AC Adapter Kit Permits operation of an XP2i-DP from an AC supply of 90 - 264 VAC and 47 - 63 Hz. Includes interchangeable international plugs (for USA, Europe, U.K., and Australia). Adapter will not charge batteries, but in the event of AC power loss, XP2is will automatically revert to battery operation. !!WARNING: Do not use the AC Adapter Kit in hazardous locations. P/N 3696 Protective Boot Skydrol™ resistant protective boot.
—— ——— — —— Support 11 Contact Us Phone ��������������������������������������������������� (805) 595-5477 Toll-Free ����������������������������������������������� (800) 444‑1850 Fax ����������������������������������������������������������� (805) 595-5466 Email ����������������������������������������������������� crystal@ametek.com Web ��������������������������������������������������������� crystalengineering.
© 2013 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 4871.