Instruction manual
14
www.cs-instruments.com
Dew point
FA 410 from -80 to 20°C
td
FA 410 – the ideal dew point sensor 
for monitoring membrane and 
adsorption dryers
Typical use in compressed air dryers/granulate dryers 
at very low dew points down to -80° C.
Special features
 Measuring range -80...20 °C
td
 Extremely long-term stable 
 Analogue output 4...20 mA
 Condensation insensitive
 Quick response time   
 Pressure-tight up to 350 bar 
  (special version)
Recommendation:
Mounting with standard measuring 
chamber for compressed air up to 16 bar 
Advantage: Easy installation via 
fast coupling.
Description      Order no.. 
FA 410 dew point sensor, -80°...20°Ctd incl. inspection certificate    0699 0410
FA 410 dew point sensor, -20°...50°Ctd incl. inspection certificate    0699 0412
Connection cables:
Connection cable, length: 5 m    0553 0104
Connection cable, length: 10 m    0553 0105
Options for FA 410:
Option special version up to 350 bar    0699 4003
Option output in mg/m
3
, g/kg, oder %RH, special scaling 4...20 mA    0699 4004
Additional accessories:
Standard measuring chamber up to 16 bar    0699 3390
High-pressure measuring chamber up to 350 bar    0699 3590
Measuring chamber for respiratory air bottles up to 350 bar    0699 3790 
CS Service Software for FA/VA 400 sensors including PC connection set,  0554 2005 
USB adapter and interface adapter to the sensor as well as 
CSM-S for data recording, please see page 21
Mains unit in wall housing 100-240 V, 10 VA, 50-60 Hz/24 VDC, 0.35 A  0554 0108 
Mains unit on DIN rail 100-240 VAC / 24 VDC, 0.35 A    0699 3340 
Calibration:
Precision calibration at -40°Ctd including ISO certificate    0699 3396
Precision calibration at 3°Ctd including ISO certificate    3200 0003
Control and calibration set 11.3 % RH    0554 0002
Control and calibration set 33 % RH    0554 0004
Control and calibration set 75.3 % RH    0554 0005
Technical data FA 410
Measuring range: -80...20°C
td
 resp. 
 -20...50°C
td
Accuracy:  ± 1°C at 20...-20°C
td
  ± 2°C at -20...-50°C
td
  ± 3°C at -50...-80°C
td
Pressure range:  -1...50 bar
special version up to 350 bar
Power supply:  24 VDC (16...30 VDC)
Protection class:   IP 65
EMC:  according to DIN EN 61326
Operating temp:  -20...70 °C
Connection:   M12, 5-pole
PC connection:  SDI interface
Analogue output:  
4...20 mA = -80...20°C
td 
resp.
4...20 mA = -20...50°C
td
Burden for
analogue output:  < 500 Ohm
Screw-in thread:  G1/2”
Dimensions:  Ø 30 mm, length approx. 
  130 mm 
Via service software:
– choose units:  % RH, °Ctd, g/m
3
, mg/m
3
, 
  ppm V/V
– scaling  change 4...20 mA
Special version:  4...20mA, 2-wire technology










