Instruction manual
FA 300 Ex -80 to 20°C
td
FA 300-2 Ex measures the dew point 
resp. the pressure dew point in 
explosive areas of Zone 1. 
Protection class: FA 300-2 Ex:
I 2G Ex ia II C T4
With the Atex approval for Zone 1 
the established dew point measuring 
instruments FA 300-2 can now be
used in explosive areas of the 
industry as well.
FA 300-2 Ex may only be used in connec-
tion with approved Ex-rated power supplies 
or safety barriers or galvanic separating 
elements 
with max.: U0 = 30 V max.
  I0 = 100 mA max.
  P0 = 1 W max.
20
www.cs-instruments.com
Dew point
Description      Order no. 
FA 300-2 Ex pressure dew point meter  0699 3070
Measuring chamber up to 350 bar  0699 3590
Special scaling  0699 4004 
Analogue output to other humidity parameters: 
% RH g/m
3
, mg/m
3
, ppm V/V, g/kg
Intrinsically safe power supply, safety barriers  0554 3071 
Special features
 Robust design
 Pressure-tight up to 300 bar
 Long-term stable humidity sensor,
  approved for years
 4...20 mA analogue output 
  in 2-wire system
 Further parameters adjustable via
  software: % RH, g/m
3
, mg/m
3
, 
  ppm V/V, g/kg
 CS Service Software for data storage 
  and calibration
  (no approval for explosive areas)
Explosive area Non-explosive area
Current supply
+ –
Intrinsically safe
power supply
L +
L –
+ +
– –
Controller/SPS
+
i=4...40 mA
–
Mains units
Description      Order no. 
Mains unit in wall housing for sensors of the Series VA/FA,  0554 0108
100-240 V, 10 VA, 50-60 Hz / 24 VDC, 0.35 A
Mains unit on DIN rail, 100-240 VAC / 24 VDC, 0.35 A  0699 3340
Power supply 100-240 VAC/24 VDC, 0.35 A for Series VA/FA 400, 2 m cable
 0554 0107
Mains unit in wall housing 왘
Mains unit in wall housing 왖
Power supply 왖
Mains unit on DIN rail 왘
Technical data 
for mains unit in wall housing
Dimensions:  118 x 133 x 92 mm 
 (WxHxD)
Power supply:  100-240 VAC, 10 VA, 
  50-60 Hz 
Output:  24 VDC, 0.35 A
Relay:  2 pieces, change-over
  contacts, 230 VAC, 3 A
Mains unit in wall housing:
Special features:
 Detachable screwing clamps
  for easy wiring 
 Galvanic separation of the pulse output
  of consumption sensors VA 400/410
 2 relay outputs (230 VAC, 3 A) 
  for alarm signal amplification at dew 
  point sensors FA 400, FA 416
 Live parts protected untouchably
  in housing
Technical data FA 300-2 Ex
Measuring range:  Pressure dew point in °C
td
FA 300-2 Ex:  -80...20°C
td
 = 4...20 mA
Pressure range:  -1...300 bar
Power supply:  24 VDC (10...30 VDC)
Accuracy:  ± 0,5°C at -10...50°C
td
  typical ± 2°C at -40°C
td
Output:  4...20 mA in 2-wire 
 technology
Protection class:  IP 65
EMC:   according to DIN EN 61326
Operating temp.:  -20...70°C
Storage temp.: -40...80°C
Burden for
analogue output:  < 500 Ohm
Screw-in thread:  G1/2” stainless steel
Housing material:  polycarbonate
Sensor protection:
Sintered fi lter 50 μm 
  stainless steel