Specifications

Table Of Contents
283
Reference Manual
00809-0100-4026, Rev HA
Appendix H: Resource block
November 2014
Resource Block
Failed
Software error
The device software has encountered an
error. Reasons may be multiple, including
too low supplied voltage, or an error
being simulated.
The status for all variables is BAD, and the
device is out of service.
1. Make sure that enough voltage is
supplied to the device.
2. Restart the device.
3. Turn off simulation of device failure.
4. If the alert persists, replace the
transmitter head.
Sensor database error
The device has found an error in the
configuration database.
Status for all variables is BAD. Device
recovery is possible.
1. Load default database to the device
to clear the error.
2. Reconfigure the device.
3. If the alert persists, replace the
transmitter head.
Internal temperature
critical
The internal temperature of the device
has reached critical levels and the
integrity of the device electronics has
been compromised. Ambient
temperature should not exceed device
specifications
(-40 °F/-40 °C to 176 °F/+80 °C).
The device is not in service (OOS) and
status for all variables is BAD.
Replace the transmitter head.
Electronics failure –
main board
The device has detected a fault with an
electrical component on the main board
electronics module assembly.
The status for all variables is BAD, and the
device is out of service.
Replace the transmitter head.
Software
incompatibility error
Fieldbus software and main firmware
versions are incompatible.
The device is not in service (OOS).
Replace the transmitter head.
Memory failure -
output board
Configuration data has been corrupted or
pending configuration changes has been
lost due to loss of power before storage
could complete.
Default values are loaded into the faulty
block. Potential errors in stored data may
cause unwanted behavior. The device is
not in service (OOS) and status for all
variables is BAD. Device recovery is
possible.
1. Load default database to the device
to clear the error.
2. Download a device configuration.
3. If the error persists, it may indicate a
faulty memory chip. Replace the
transmitter head.
Internal
communication failure
The communication between the main
transmitter board and the fieldbus
electronics board has been lost.
The device holds the last known
measurement values with status BAD.
Replace the transmitter head.
Electronics failure -
output board
The device has detected a fault with an
electrical component on the output
board electronics module assembly.
The device is not in service (OOS).
Replace the transmitter head.
Alarm type
Failed/maint/advise/
active event
Description Recommended action