3.3.2 Special considerations in solids applications
The transmitter should be mounted as close to the center of the tank as possible, but
not in the center of the tank. A general practice is to mount the transmitter at
2
/
3
tank
radius from the tank wall, see Figure 3-4.
Figure 3-4. Transmitter Location in Solids Applications
The radar signal must never be shaded by the inlet nor the injected product, see
Figure 3-5.
Figure 3-5. Install the Transmitter with a Clear and Unobstructed View