How-To Guide

FUSE REPLACEMENT
FUSE TEST
WARNING
WARNING
Always perform a Fuse Test before using the
internal current shunt measurement mode.
Only use appropriately rated replacement fuse.
Open the N2 - Neuron per directions above.
Gently remove existing fuse from holder.
Replace fuse only with appropriately rated replacement.
Recommended model:
1.
2.
3.
4.
5.
Reomax 632.300.12, available from Curien LLC
Size 3AG 12A hold 600V 10kA minimum interrupt
Close the N2 - Neuron per directions above.
Perform a resistance measurement with the C and Ω
terminals connected
1.
2.
If the resistance measured is greater than 0.5Ω, replace
the fuse*
* The resistance mode on the N2 - Neuron is not accurate below
1Ω. The actual resistance of the fuse should be 20mΩ.
N
2
FIGURE 8 FUSE TEST