Use Instructions

45
Implantable Pulse Generator Parameters
Parameter
Description
Units
Programmable Values
Therapy
Schedule
From/To Times for Therapy (N) or
Therapy Off
HH:MM
Up to 3 entries allowed
Any time during the day
In 15 minute steps
Pulse
Amplitude
for Therapy
(N)
The amplitude of each applied pulse.
milliamp
1.0 to 20.0
Pulse Width
for Therapy
(N)
The width of each applied pulse.
µs
15 to 500
Therapy
Frequency
for Therapy
(N)
The frequency of applied pulses except
during the Rest portion of the Burst
Interval.
PPS
10 to 100
Burst
Not Checked = therapy pulses are applied
throughout the burst cycle in a
continuous manner
Checked = pulses are applied in a cycle
of active and rest periods.
N/A
Not checked /
Checked
Burst Duration
The length of the active portion of the
burst cycle during which the Therapy
Frequency is delivered.
NOTE: This parameter is not shown if
Burst is not checked.
milliseconds
50 to 1950
Burst Interval
The total length of the burst cycle
including the active portion and the rest
portion.
NOTE: This parameter is not shown if
Burst is not checked.
milliseconds
100 to 2000
Implantable Pulse Generator Longevity
The battery lifetime of the IPG is dependent on device therapy settings. Assuming 825 Ohm lead
impedance the following table indicates the resulting longevity based on different therapy settings.
For these calculations, a single 24-hour therapy was assumed.
Pulse
Amplitude
(mA)
Pulse Width
(us)
Therapy
Frequency
(Hz)
2104 Device
Longevity
(Months)
2102 Device
Longevity
(Months)
4.2
125
40
100
79
5.6
125
40
74
60
7.2
125
40
55
44
*8.0
250
40
25
28
*Worst case conditions