Use Instructions

50
Programmer Miscellaneous Information
Description
Information
Type of protection against electric
shock
The Programmer Interface is not mains powered
equipment.
Degree of protection against electric
shock
The Programmer Interface meets IEC 60601-1 touch
current requirements.
Degree of protection against the
ingress of water
Ordinary
Methods of sterilization or disinfecting
Cannot be sterilized.
Information regarding electromagnetic
or other interference and advice
regarding avoidance as necessary.
Do not use in the proximity of equipment that generates
electromagnetic interference (EMI). EMI may cause a
disruption in programmer function. Examples are cell
phones, x-ray equipment, and other monitoring
equipment.
Accessories or materials used with
equipment that may affect safety.
USB cable to connect computer/tablet to Programmer
Interface
Cleaning and maintenance, with
frequency
If the Programmer System requires cleaning, clean the
system components with a soft cloth dampened with
water. Do not allow pooling or ingress of liquid into the
Programmer Interface enclosure.
No preventative maintenance is required.
Do not use programmer system if programming unit or
cables appear damaged.
There are no serviceable items.
Please contact CVRx
®
representative to return product
for service or replacement.
Equipment Supply Disconnect
Unplug power cord to isolate equipment from supply
mains.
Manufacturer Name
CVRx, Inc.
Model #(s)
Programmer System: Model 9010
Programmer System Model 9020
Power Supply
9010:
Input Voltage: 100-240V
Input Current: 0.6A
Input Frequency: 50/60Hz
Output Voltage: 20V
Output Current: 3.25A
Output Power: 65W
9020:
Input Voltage: 100-240V
Input Current: 0.6A
Input Frequency: 50/60Hz
Output Voltage: 15V
Output Current: 1.6A
Output Power: 24W