Use Instructions

61
PROGRAMMER SYSTEM EMC PRECAUTIONS
The Programmer System requires special precautions regarding Electromagnetic
Compatibility (EMC) and is to be installed and put into service according to the EMC
information provided in this guide.
Portable and mobile RF communications equipment can affect the Programmer System.
The use of power cords or USB cables other than those supplied with the Programmer
System may result in increased emissions or decreased immunity.
The Programmer System should not be used adjacent to or stacked with other equipment.
If such use is required, then the Programmer System should be observed to verify normal
operation in this configuration.
PROGRAMMER SYSTEM RF SPECIFICATIONS
The Programmer System may be interfered with by other equipment, even if that other
equipment complies with CISPR emission requirements. The RF telemetry operating
specifications are:
MICS band 402-405 MHz. The effective radiated power is below the limits specified in:
USA: 47 CFR 95 Subpart I
Canada: RSS-243
2.4 GHz band 2.4-2.4835 GHz. The effective radiated power is below the limits specified
in:
USA: 47 CFR 15.249
Canada: RSS-210