Use Instructions

66
Table 4: Separation Distance
Recommended separation distance between portable and mobile RF communications equipment and
the Programmer System
The Programmer System is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Programmer System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Programmer System as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
Pd
=
3
5,3
80 MHz to 800 MHz
Pd
=
3
5,3
800 MHz to 2,5 GHz
Pd
=
3
7
0,01
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
1,2
1,2
2,3
10
3,7
3,7
7,4
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.