User's Manual

2/2020 CY-0090-018, Rev.A Page 42 of 49
Electrical fast
transient/burst
IEC
61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not Applicable
Battery powered
Not Applicable
signal lines less
than 3 meters
Mains power quality should be
that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to
earth
Not Applicable
Battery powered
Mains power quality should be
that
of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on
power supply
input
lines
IEC
61000-4-11
<5% U
T
(>95% dip
in U
T
for 0.5 cycle
40% U
T
(60% dip
in U
T
)
for 5 cycles
70% U
T
(30% dip
in U
T
)
for 25 cycles
<5% U
T
(>95% dip
in U
T
)
for 5 s
Not Applicable
Battery powered
Mains power quality should be
that of
a typical commercial or hospital
environment. If the user of the e-vive
requires continued
operation during
power mains interruptions, it is
recommended
that the e-vive
be
powered
from an uninterruptible
power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic
field
IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a
typical commercial
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
T