Operating instructions

3-32 (E)
DFP-R3000
3-6. FG checking
Perform the FG checking when sprocket SUB assembly,
sprocket holder and brake are replaced.
Required Equipment and Tools
DFP-D3000
Projector
Oscilloscope
Connection cable (Supplied with DFP-R3000)
Alignment film (with data sheet)
Preparation
Remove the front panel assembly and left side panel and
then connect the DFP-R3000 and DFP-D3000 READER
connector using connection cable.
Procedure
(1) Connect the oscilloscope to the following points on the
RD-35 board.
Oscilloscope RD-35 board
CH1 TP4 (FG)
(2) Load the alignment film into the DFP-R3000 and
projector, and then run the film.
(3) Check that the output (FG waveform) of oscilloscope
meets the following specification.
Specification : Duty ratio
B
= 50 ±10 %
A
(4) Change the measuring range of the oscilloscope as
shown below, and check that the FG waveform of the
oscilloscope is not chipped.
(5) If the duty ratio does not meet the specification,
perform the installation of the replaced part again.
3-6. FG checking