Product Info

Page 19/26
3.4.1. What to do if the WORK AIR DL001 T U is triggered
When triggered, the system will inflate and then deflate slowly. The airbag can only inflate
once, After each deployment the WORK AIR DL001 T U must be replaced; contact the
DAirlab Assistance Service for the replacement procedure.
WARNING:
●●● in the event of a fall during which the System is triggered, do not use it again until
the airbag has been replaced by the DAirlab assistance service, as the System will
not function.
3.5. Guide to the LED sequences during use
The user interface consists of a multicolor illuminated button and a vibramotor integrated in
the printed circuit board located in the front left part of the System.
3.5.1. System operating mode
The system has the following operating modes:
“Init Mode” (Initialization mode): airbag protection is not available.
“Normal Mode”:
Typical mode of use of the system, airbag protection is available. When the battery
level is below 25%, airbag protection is available, the residual charge is less than 2
hours of continual use. Charge the battery as soon as possible.
“Error Mode”: Mode of use when the diagnostic system detects an error involving
the electronic components integrated in the main board or involving the state of the
gas generators integrated in the airbag.
WARNING:
●● In the event of an error, switch off, check the status of the connections and then switch
the System on again. If the error persists when the System is switched off and on
consecutively more than three, it will always start in “Error Mode”. This might indicate a
System malfunction, in which case contact the DAirlab Assistance Service.
“Fired Mode”: deployed mode following triggering of the airbag
“Charging Mode”: mode activated during charging.