Appendix A | Specifications
Code Numbers S 2450 S 2451
Electrical 100-120 VAC @ 3.0 A 220-240 VAC @ 1.6A
Processing Time 0-99 hours
Program Number 40
Heating Time 37°C to 95°C; 2 minutes
Cooling Time 95°C to 45°C; 5 minutes
Program Types
Denaturation and Hybridization
Hybridization only
Fixed temperature
Temp. Uniformity Within +/- 1°C on the heat plate
Capacity Up to 12 slides
Dimensions Depth 45.1 cm/ 17 3/4 in
Width 22.8 cm/ 9 in
Height 14.0 cm/ 5 1/2 in
Weight 8.5 kg
Environmental Indoor use
Altitude up to 2000m
Temperature 15°C to 40°C
Maximum relative humidity 80% for temperatures up to 15°C decreasing
linearly to 50% relative humidity at 40°C
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Pollution degree 2
Dako Warranty
Dako warrants to Buyer that the Hybridizer will be free, under normal use and maintenance, from defects
in material and workmanship for a period of one (1) year from the date of delivery to the end-user. As
Buyer's sole and exclusive remedy for any such defects, Dako will, at its option, repair or replace without
charge any unit which is found to be so defective.
Limitation of Liability
Dako shall not be responsible under this warranty for Products found to be defective as a result of
misuse, mishandling, or neglect.
DAKO MAKES NO OTHER WARRANTIES OTHER THAN THAT SPECIFIED ABOVE, OF ANY KIND
WHATSOEVER, EXPRESS OR IMPLIED, AND ALL WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED BY DAKO AND EXCLUDED
FROM THE TERMS OF SALE.