Bayer Raw8 used during calibration (left) – RGB 32-bit, UYVY, or YUY2 for image acquisition (right)
Start the Flat Field calibration tool via the CamExpert menu bar:
Pre-processing • Flat Field Correction • Calibration.
Flat Field Calibration Window
The Flat Field calibration window provides a three step process to acquire two reference images and then save the
flat field correction data for the Genie used. To aid in determining if the reference images are valid, a histogram tool
is provided so that the user can review the images used for the correction data.
CamExpert Flat Field Calibration Menu
• Click on the Advanced Setting button to change the default number of frames averaged for each calibration
step. The default value is 10 frames.
• Setup the camera to capture a uniform black image. Black paper with no illumination and the camera lens’ iris
closed to minimum can provide such a black image.