Differential Connections for Ground-Referenced
Signal Sources
Figure 4-4 shows how to connect a ground-referenced signal source to
a channel on the device configured in DIFF input mode.
Figure 4-4. Differential Input Connections for Ground-Referenced Signals
With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V
cm
in Figure 4-4.