User manual
Chapter 3 Software Overview
© National Instruments Corporation 3-3 NI-DAQ User Manual for PC Compatibles
Set_DAQ_Device_Info
Selects parameters pertaining to the device
operation.
Timeout_Config
Establishes a timeout limit that is used by
the synchronous functions to ensure that
these functions eventually return control to
your application. Examples of synchronous
functions are
DAQ_Op
,
DAQ_DB_Transfer
,and
WFM_from_Disk
.
Software-Calibration and Device-Specific Functions
Each of these software-calibration and configuration functions is specific
to only one type of device or class of devices.
Refer to the NI-DAQ Function Reference Online Help file to determine
which functions your device supports.
AO_Calibrate
Loads a set of calibration constants into
the calibration DACs or copies a set of
calibration constants from one of four
EEPROM areas to EEPROM area 1. You
can load an existing set of calibration
constants into the calibration DACs from
a storage area in the onboard EEPROM.
You can copy EEPROM storage areas
2 through 5 (EEPROM area 5 contains the
factory-calibration constants) to storage
area 1. NI-DAQ automatically loads the
calibration constants stored in EEPROM
area 1 the first time a function pertaining to
the AT-AO-6/10 is called.
Calibrate_1200
Calibrates the gain and offset values for
the 1200/AI devices ADCs and DACs. You
can perform a new calibration or use an
existing set of calibration constants by
copying the constants from their storage
location in the onboard EEPROM. You can
store up to six sets of calibration constants.