User Manual

WMTS TRANSCEIVERS
198 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
Remote Record
When depressed, the Remote Record function button will initiate a strip chart
recording at the Central Station.
Nurse Call
When depressed, the Nurse Call function button will initiate a Nurse Call Alarm at the
Central Station.
Power Button
Pressing the Power button either places the transceiver in or out of Standby Mode.
Attendant Present / Procedure Alarm Silence (PAS) Unlock Button
The Attendant Present push button has two functions. Each function is initiated based
on how long the button is pressed.
Note: The LEDs are constantly illuminated when powered by an external source.
1. Initiating an Attendant Present Alarm
Once the transceiver is out of the Standby Mode, pressing the Attendant Present but-
ton will activate the Attendant Present function and initiate an Attendant Present
Alarm at the Central Station.
2. Unlocking the PAS button
The PAS function must be enabled at the Central Station prior to initiating the Proce-
dure Alarm Silence alarm at the DT-7000/DT-7001.
In the “locked” position, the PAS button is disabled.
To “unlock” the PAS button, press, and hold (for about two seconds), the Attendant
Present button until the Procedure Alarm Silence Status Indicator LED begins flash-
ing. Once the LED indicator starts flashing, the PAS button is in the “unlocked mode”
and functional.
Note: The PAS button must be pressed while the LED is still flashing. If it is
pressed after the LED has stopped flashing, then the PAS button will auto-
matically be “re-locked”.