PATIENT SETTINGS
88 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
4. Press the Bottom Wave button to select the non-ECG wave. The choices are
PLETH, P1, P2, RESP, CO2 and None.
5. Select the size of the bottom wave by pressing Bottom Size. Choices are as
follows:
P1 and P2: 1.5, 3, 6, 12 or 24 mm/Hg, where 24 is the smallest and 1.5 the
largest.
RESP, CO2 and PLETH: 0.25x (smallest), 0.5x, 1x, 2x, 4x (largest).
Labeling Waveforms
You can select labels for each transmitted wave; the number of available waveforms
depends on the device type selected in the Patient Settings screen. Available wave-
forms are shown in table 9. OpenNet-compatible bedside monitors can transmit up to
three waves of data.
To label the waves:
1. Press Label Set on the Patient Settings screen to display the Label Set Defini-
tion popup shown in fig. 38. See table 9 for available wave labels.
Fig. 38. Label Set Definition popup
2. Press the three Wave buttons to cycle through available waves.
3. For 4- and 5-wire transceivers, toggle the Augment button to Yes if you wish
the system to derive the augmented leads aVR, aVL, and aVF automatically.
If augmented leads cannot be derived from the leads, N/A appears.
Note: Lead III is derived regardless of the Augment setting.
4. Press Exit on the popup to save the new waveform labels.