VIEWING CURRENT PATIENT STATUS
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 133
All information contained herein is subject to the rights and restrictions on the title page.
Automatic Alarm Printing
You can configure the system to print alarmed events automatically when they occur.
To do so, you must enable individual alarms for automatic printing as instructed
below.
1. Press Setup on the Main screen and select the patient’s waveform.
2. Press Alarm Config on the Patient Settings screen to display the Alarm Con-
fig screen.
3. Highlight the desired alarm and toggle Record to On. This alarm will now
trigger automatic printouts.
4. Repeat this step for all alarms you wish to print automatically for this patient.
You can also print manually at any time.
Printing at the PatientNet Viewers
All recording requests made at a PatientNet Viewer (also known as an IRVS or RVS)
are recorded at the IRVS/RVS.
All printing requests made at the IRVS/RVS are printed at the same IRVS/RVS. Full
Disclosure strips can be printed at either the Central Station, PatientNet Viewer, or the
Interactive-PatientNet Viewer.
Printing from the Bedside Monitor
When you request a printout from the bedside monitor and DT-7000 instrument trans-
ceiver notifies the Central Station, which generates a real-time strip for the corre-
sponding patient channel as if the record button had been pressed.
If the alarm is set to Record and Store on the Alarm Config screen, the real-time strip
is printed and stored in history.
Printing Real-time Waveform Strips For All Patients
You can print real-time waveforms for all admitted patients displayed on the Central
Station at one time:
1. Press System on the Main screen.
2. Press Record All button on the Passcode screen. A real-time waveform strip
is printed to the laser printer or the strip recorder for all patients on the central
Station.