Changing Survey Page Configuration
Prior to first use in a care unit, the effective range and coverage
area of each transmitter should be tested by conducting a paging
survey to ensure that the transmitter and pagers are performing
as needed.
The programmer can be used to set the time between pages and
the total number of pages to be sent for performing survey
paging (the default in 100 pages, sent every 15 seconds).
To set interval between pages and total number of pages:
1. From SysMgr, tap Start button and tap Install Program.
2. Enter password. Alarms screen will be displayed.
3. Tap Monitor tab to display Monitor Configuration screen.
{Insert monitor configuration screen (Figure 6-9)}
Monitor Configuration Screen
!!
!!
!!
Monitor Type and Authorization fields are read-only values set at
initial installation. You will not need to edit these values.
4. To change survey page time intervals, tap Down Arrow in
the sec (seconds) field. Tap to select desired time interval
from pull-down menu.
5. To change total number of survey pages, tap Down Arrow
in the times field. Tap to select desired number of pages
from pull-down menu.
6. Program survey page configuration into transmitter. Turn
transmitter on and aim infrared port on programmer directly
at infrared port on front of transmitter and tap SEND button.
Programmer will display a “Sending Program Data” status
message.