Service Manual

AlarmView™ System User’s Guide
{insert part #} Page 33
4. Tap to select desired time interval from pull-down menu.
5. Aim programmer infrared port directly at infrared port on
front of transmitter and tap SEND to program configuration
changes into transmitter. A “Programming OK” message will
be sent to pager(s) assigned to transmitter.
6. Repeat programming for every transmitter in system.
7. Tap Exit button to return to SysMgr.
Section 5: Testing and Periodic
Maintenance
All tests described in this section can be performed without
disassembling the units in the AlarmView System. Tests should be
conducted prior to initial installation, when adding a new unit to the
system, and before returning equipment to service after repairs or routine
maintenance.
!
!
!
If any component of the system fails to perform as specified, do not
return it to use. Contact Data Critical Corp. regarding repair or
replacement of the equipment.
Testing System Safety
AlarmView safety tests meet the standards of, and are to be performed
in accordance with, CAN/CSA-C22.2 No. 601.1 and UL 2601-1, Medical
Electrical Equipment Part 1: General Requirements for Safety, for
instruments classified as Class 1 and Not Classified.
!!
!!
!!
Equipment Needed is a Safety Analyzer that must meet ,
CAN/CSA-C22.2 No. 601.1 and UL 2601-1 specifications. Bio-
Tek Instruments, Inc. or compatible.