Service Manual

AlarmView™ System User’s Guide
{insert part #} Page 35
Communications Page
Each time a transmitter is turned on, or re-establishes
communication with a monitor after having lost communication,
the transmitter will automatically send a communications page
(Comm OK Page) to all pager(s) assigned to the transmitter.
After setting up transmitter (connecting it to power and to a
monitor), to send a Comm OK Page test:
1. Check monitor to ensure it is on and functioning properly.
2. Press On/Off button on top of transmitter.
Indicators will flash on and off while transmitter performs a
Power-On Self Test (POST) to check its function.
Mode LED is solid green when transmitter has established
communication with monitor.
Tx LED will flash green, and transmitter will send a Comm
OK Page to assigned pager(s) to confirm communication
between transmitter and monitor. The message will display:
Bed: 1234
Comm OK
All Pagers*
*Last line All Pagers with be replaced with “Pager 1234” if one
pager is assigned to transmitter, and “Pager 1234 5678.” The 4
digits correspond to the last 4 digits of the assigned pager(s)’
cap code.
3. Verify that page is sent to assigned pager(s) and that bed
number is correct.
The transmitter will now send pages to the assigned pager(s)
whenever it detects alarms identified by the monitor.