Service Manual

AlarmView™ System User’s Guide
{insert part #} Page 36
Test Page
You should test the AlarmView System setup by conducting a
Test Page when any changes affecting the system have
occurred.
To perform a Test Page, the transmitter must be on and
communicating with the monitor.
To send a test page:
1. Press On/Off button to turn transmitter on. Check to ensure
that transmitter is communicating with monitor (both AC
Indicator and Communications Indicator on transmitter
are Green).
2. Turn on pager(s) assigned to transmitter.
3. To send a test page, press and release On/Off button again.
All pagers assigned to the transmitter will display the
following page (if a patient is not connected to the sensor,
dashes will be displayed for SpO2 and pulse rate):
Bed: 1234
Test Page
SPO2: 83%
PR: 70 bpm*
* For the Nellcor N-3000 monitor, the last line of the test page will display
heart rate (HR).