User Manual

AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 26
Turning Transmitter On
The antenna must be connected to the transmitter before the
transmitter is turned on. Turning the transmitter on without the
antenna connected may damage the transmitter. Do not disconnect
the antenna while the transmitter is turned on.
1. Check oximetry monitor to ensure it is plugged in (not
operating on battery), functioning properly, and that alarm
settings are appropriate.
2. Press On/Off button on top of transmitter.
Indicators will flash on and off while transmitter performs a
Power-On Self Test (POST) to check its function.
Transmitter LED Indicators
POST Light Sequence
Stage Stage Indication Test Failure Indication
1 Green
Transmitting LED
Transmitter
program flash
memory
check
Simultaneous flashing
of Tx and Mode LEDs
2 Green
Communications LED
External
SRAM
Simultaneous flashing
of Tx and Mode LEDs
3 Amber Operation of Simultaneous flashing