AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 40
3. Tap Exit button (< in upper right corner) to return to SysMgr.
Changing Survey Page Configuration
Prior to its first use in a care unit, the effective range and
coverage area of each transmitter should be tested by
conducting a paging survey to ensure that the transmitter and
pagers are performing as needed.
The programmer can be used to set the time between pages and
the total number of pages to be sent for performing survey
paging (the default in 100 pages, sent every 15 seconds).
To set interval between pages and total number of pages:
1. Tap the Start button and tap the AlarmView Install
Program.
2. Enter your password. The Alarms screen will be displayed.
3. Tap the Monitor tab to display the Monitor Configuration
screen.
Monitor Configuration Screen