User Manual

AlarmView™ System Administrator’s Manual
880-0326-01 REV. A Page 61
Unable to Access
Port Probably in Use
Programmer error. Reboot programmer by
inserting stylus in the
reset button located on
the back on the
programmer (see
appendix X).
Contact Data Critical
regarding repair and/or
replacement of
programmer.
You must select or
get a bed number and
one or more pagers
Programmer SEND
button was pressed
before selecting a
bed number or pager
numbers.
Follow procedure to
select bed and/or pager
numbers and tap SEND
again.
Missing pages
Indication Possible Cause Corrective Action
"Programming OK"
page was not
received at the
assigned pager(s)
after programming
the transmitter.
The pager is
incorrectly assigned
to the transmitter.
The pager's battery is
too low or depleted.
The pager is not
turned on.
Pager defective.
Transmitter
Defective.
Follow procedure to
select bed and/or pager
numbers, ensure that
the correct pager is
selected.
Replace pager's
battery.
Turn on pager.
Contact Data Critical
Corp. regarding repair
and/or replacement of
transmitter or pager.