User's Manual

26SoftwarePhysioTelDigitalConfigurationManual
The Enabled mode allows the software system to record, store, and analyze data from the implant.
Ì WARNING: If the implant is notEnabled” the device will still be in communication with the
system but no data from this implant will be visible to the analysis software.
Current Mode:
The Turn Implant Off button allows the user to remotely switch the implant to the OFF mode. Once in the
OFF mode, the implant cannot be remotely returned to the ON mode. The implant can only be turned ON by
physically passing a strong magnet close to the device for a few seconds.
Ì WARNING: Once the implant is switched OFF, it cannot be remotely turned back ON.
Procedure for switching the implant to the OFF mode:
1. Select the implant by clicking the line item in the Configuration column on the left of the screen.
2. Click the button labeled Turn Implant Off.