User's Manual

46SoftwarePhysioTelDigitalConfigurationManual
Reset
The Reset function returns the TRX settings to the factory default values.
1. Select Reset from the TRX Command: drop-down menu.
2. Select the TRX # from the TRX No: drop-down menu.
3. Change the value in the dialog box below the letters TRX from “ff” to “02”.
4. Click the Send button.
5. A successful operation is indicated by a blue colored Command Completed banner at the top of the
screen and a text string “OK” below the word Reply at the bottom of the screen.
TRX RSSI History
This option allows the user to sample how well the TRXs are receiving RF signals from the implants. In an
actively running system these graphs continually update according to a user prescribed auto refresh rate.
There will be one RSSI graph displayed for each of the enabled TRXs connected to the CLC. The TRXs will
average the received signals from all of the implants it is communicating with. It will then average the
implant signals and report an average value for each graph.
Follow this procedure to utilize the TRX RSSI History option:
1. Click on the Settings
link under the CLC Options heading.
2. In the text box to the right of the words RF Mode, delete the word “Enabled
3. Type the word “Assessment” in the text box and press click the Save button below.
4. Using the mouse cursor, click on the link marked TRX RSSI History.