User's Manual

50SoftwarePhysioTelDigitalConfigurationManual
o Standby – not actively transmitting data
o Active – actively transmitting data
o Unused – configured but probably out of range
Next Mode: (This feature is not yet implemented).
Implant Commands
There are three commands with which the user can communicate with individual implants. They are Ping;
Get RSSI Threshold; and Set RSSI Threshold.
The Ping command allows the user to select an individual implant and request a confirmation
message that the implant is operating within range.
The Get RSSI Threshold command retrieves the current threshold value from the implant.
The Set RSSI Threshold command allows the user to adjust the lower limit of signal strength that
the implant will accept as viable information from any of the TRXs.
The “Ping” Command
The Ping command allows the user to send a request to an individual implant to reply with a confirmation
message that the implant is operating within range.
Click on the drop-down menu labeled Implant ID:
Select a device by left clicking on an implant serial number.