User's Manual

54SoftwarePhysioTelDigitalConfigurationManual
Common Error Codes
The implant commands in this section are capable of generating an error code if the command cannot be
successfully executed. Below is a list of the more common error codes.
Error code Description Solution
900
Unknown Error
901
Implant Not Found Make sure implant in range
902
Timeout Make sure implant in range
903
Send Fail Make sure implant in range, in standby mode, Ponemah
is not trying to send a lot of commands to the implant
905
Implant in Active Mode Make sure the implant is in standby mode
906
Queue Full Make sure implant is in range, Ponemah is not trying to
send a lot of commands to the implant
Implant RSSI History
Similar to the TRX RSSI History, the Implant RSSI History generates graphs in which the received signal
strength from each of the implants is plotted for each implant. These graphs allow the user to track how well
the implants are being received by each of the TRXs. In an actively running system these graphs continually
update according to a user prescribed auto refresh rate.
There will be one RSSI graph for each of the recognized implants in the system.
Each implant will report the received signal strength from each of the TRXs it is communicating with.
The RSSI graph will display one data set for each of the TRXs.