User's Manual

TruSense Configuration Manual TruSense Hardware Configuration 41
3. A successful operation is indicated by a blue colored Command Completed banner at the top of the
screen and a text string below the word Reply at the bottom of the screen.
4. The reported text value OK “xx” is the Hex value of the RSSI Threshold.
Set RSSI Threshold
RSSI stands for Received Signal Strength Indicator. It is a quantitative measure of the strength of the RF
signal that the TRX is receiving from the implants. The Set RSSI Threshold command allows the user to
adjust the lower limit of signal strength that the TRX will accept as viable information from the Implants.
The default value = 0x12
NOTE: Anytime the TRX is unplugged, or the CLC is rebooted, the RSSI threshold value will revert
back to the default value of 0x12.
1. Select Set RSSI Threshold from the TRX Command: drop-dowmn menu.
2. Select the TRX # from the TRX No: drop-down menu.
3. Enter a value in the small text box above the Send button.