User's Manual

50 TruSense Hardware Configuration TruSense Configuration Manual
6. A successful operation is indicated by a blue colored OK banner at the top of the screen.
7. A Hex value will also be reported in a table below the word Reply.
Set RSSI Threshold
RSSI stands for Received Signal Strength Indicator. It is a quantitative measure of the strength of the RF
signal that the TRX is receiving from the implants. The Set RSSI Threshold command allows the user to
adjust the lower limit of signal strength that the TRX will accept as viable information from the Implants.
1. Click on the drop-down menu labeled Implant ID:
2. Select a device by left clicking on an implant serial number.
3. Click on the drop-down menu labeled Implant Command.
4. Left click the Set RSSI Threshold command.
5. A small text-entry box will appear below the Implant Command: line.
Allowable options for RSSI Threshold
0C
1C
6. Enter a new value for the RSSI Threshold and click the Send button
7. A blue colored banner with the word OK! will appear at the top of the screen.