INTRODUCTION 
2
7. 
Configure the other device parameters via VisiSet™ [Operating Mode, Calibration, Data 
Collection parameters, etc.]. 
8.  If using the CBX connection box equipped with a BM100 Backup module, perform Device 
Backup at the Slave. 
The Slave device is now Configured. Repeat these steps for each Slave reader in the ID-
NET™ network. 
Configuration of Master
1.  Connect a PC equipped with VisiSet™ to the Main, or Auxiliary interface of the planned 
Master reader. 
2.  Launch VisiSet™ and connect to the Master reader. 
3.  From the VisiSet™ Device Menu select "Parameter Setup". 
4. Set the Role of the Master reader (Synchronized or Multidata) from the 
Reading System Layout > Device Network Setting > Topology Role parameter. 
5. Enable the planned Slave device N from the Reading System Layout > 
Expected Slave Device #N > Status parameter and, if desired, set the related 
identification string from the Expected Slave Device #N > Device Description parameter. 
Repeat this step for all planned Slave devices. 
6. If necessary, set the ID-NET™ baudrate from the Reading System Layout > 
Device Network Setting > Network Baud Rate parameter, (500 kbs default). 
7.  Configure the other device parameters via VisiSet™ [Operating Mode, Calibration, Data 
Collection parameters, etc.]. 
8.  If using the CBX connection box equipped with a BM100 Backup module, perform Device 
Backup at the Master. 
9.  From the VisiSet™ Device Menu select "ID-NET™ Status Window" and click on the 
"Look For Devices On Network" button to check the status of the expected Slave devices 
within the ID-NET™ network. 
The reader network is ready. 