Motion Features
Motionless Timeout
This setting specifies the amount of time that the reader takes to assume that it is in a motionless
condition. The range for this setting is from 500 msec to 25.5 seconds, in 100 millisecond
increments.
Follow these instructions to set this feature.
1. Determine the desired setting.
2. Pad the result with leading zeroes to yield three digits. For example: 0.5 = 0005 = 00, 5 =
0
50,
20 = 200, etc.
3. Go to page 254 and scan the ENTER/EXIT PROGRAMMING MODE bar code to
enter Programming Mode.
4. Scan the bar code: SELECT MOTIONLESS TIMEOUT SETTING.
5. Scan the appropriate three digits from the keypad in Appendix D, Keypad representing
the duration which was determined in the steps above. You will hear a two-beep indica-
tion after the last character.
If you make a mistake before the last character, scan the CANCEL bar code
to abort and not save the entry string. You can then start again at the
beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE bar code to exit Programming
Mode.
This completes the procedure. See Tabl e 48 for examples of how to set this feature.
Table 48. Motionless Timeout Setting Examples
STEP ACTION
EXAMPLES
1 Desired Setting
500ms 1 sec. 10 sec 9,900ms (9.9 sec.)
2
Divide by 100 (and pad with
leading zer
oes to yield two
digits)
005 010 100 250
3 Scan ENTER/EXIT PROGRAMMING MODE
4 Scan SELECT MOTIONLESS TIMEOUT SETTING
5
Scan Two Characters From
Appendix D, Keypad
‘0’, ‘0’ and ‘5’ ‘0’, ‘1’ and ‘0’ 1 ’, 0 ’, a n d 0 ‘2’, ‘5’, and ‘0’
6 Scan ENTER/EXIT PROGRAMMING MODE