7. Re-wipe the surface of the meter lens with a clean, new cotton swab each time until
the swab comes up clean, and the clean lens squeaks when wiped.
8. After cleaning, ensure that the lens is completely dry; allow time for the alcohol to
evaporate
Background Check for Firmware Version 0210 and Prior
Sensor Firmware Versions Only
1. Connect the ROADRUNNER to the FIREFLY via the PPU.
2. On the ROADRUNNER 860 Press
SHIFT I to display the FIREFLY main menu
and on the ROADRUNNER X7 press F5
.
3. For 0210 and prior sensor firmware versions, Press
A to begin testing the FIREFLY
optics. For 0211 and future sensor firmware versions go to the Sensor Placement
section below.
Verifying
Sensor Placement
Background: 181
Status: [Success]
Test Again?
Y/N
The values shown are in decimal and the test keys on Power Level 2 (P2). A successful
test and sensor target placement is achieved when the P2 value is 60 or greater.
*After completing the sensor background test, determine if another alcohol cleaning is
needed on the register face prior to installation of the sensor.
Sensor Flap
Within each case of FIREFLYs, 24 rubber Sensor Flaps are included for every
installation. Utilizing these Sensor Flaps insures a stable environment during the
calibration process.
Sensor Placement
1. Insert sensor cable through sensor flap.
2. Remove the adhesive backing from the high-bond tape on the Optic Sensor face.
3. Install sensor to the location of the best background according to the background
check (for 0210 and prior sensor firmware version FIREFLYs).