Operator`s manual
78  A-Class 
Standard Calibration (continued) 
 Note: Do not move the position of the Media Sensor after it has been adjusted. 
Step  Action  Displayed Message  Comment 
9 
Observe the display. 
GAP MODE 
CALIBRATION COMPLETE
- Or, for reflective media: 
!
REFLECTIVE MODE 
CALIBRATION COMPLETE 
- Or, for continuous media: 
CONTINUOUS MODE 
CALIBRATION COMPLETE 
!!
!
The calibration was successful. 
 Note: If ‘Warning Low Backing’ 
is displayed (most typical with 
notched media or media on a 
transparent liner), the message 
indicates that calibration was 
successful. For other possible 
messages see Section 6.2. 
10 
Press the 
ESC
 Key to exit the 
menu. 
If using gap or reflective 
media, press and hold the 
FEED
 Key. 
READY 
!
Allow the printer to output at 
least two labels.  
The printer is ready to begin 
printing. 