Operator`s manual
I-Class  69 
Advanced Entry Calibration (continued) 
Step
Action  Displayed Message  Comment 
14 
a) Place the media in the 
sensor. Record the 
reading and label it “P” 
(paper). 
b) Place the backing, notch, 
or mark in the sensor. 
Record the reading and 
label it “G” or “M” (Gap or 
Mark). 
c) Remove media. Record 
the reading and label it 
“E” (Empty). 
GAIN TRAN  <yyy> 
*08  <0 - 31> 
Or, for reflective media: 
GAIN REFL  <yyy> 
*08  <0 - 31>
The samplings using the 
selected Gain Number 
are taken. 
15 
Press the 
 Key, and 
then press the 
 Key. 
ADVANCED ENTRY 
SENSOR LEVELS
The sensor readings 
must be entered into the 
printer. 
16 
Press the 
 Key. 
Using the 
 Key or the 
 Key, set the Paper 
Sensor Level to the value 
determined in the previous 
step. 
Press the 
Key to 
select the entry and 
advance the menu. 
Repeat for the Gap Sensor 
Level (or Mark Sensor Level) 
and the Empty Sensor Level 
entries.
PAPER SENSOR LEVEL 
P*198 G*084 E*014 
 
GAP SENSOR LEVEL 
P*198 G*084 E*014 
 
EMPTY SENSOR LEVEL 
P*198 G*084 E*014 
Or, for reflective media: 
PAPER SENSOR LEVEL
P*015 M*181 E*213 
 
MARK SENSOR LEVEL
P*015 M*181 E*213 
 
EMPTY SENSOR LEVEL
P*015 M*181 E*213
Selection is denoted by 
an asterisk (*). 
The displayed message 
will change and the 
selection will flash to 
indicate the next entry. 