Instructions

UM-0085-B09 DT80 Range User Manual Page 119
RG
The following "screens" will then be available. These can be scrolled through using the Up and Down arrows on the
keypad. Pressing Down
Display screen
Comments
DT80 V9.00
MYJOB
The default "sign-on" screen indicates the DT80's firmware version number and the name of the
currently loaded job (No current job is displayed if there isn't one)
Date: 23/10/2009
Time: 16:44:02
Current date and time (format can be changed using P31 and P39)
Battery: 90% ↓
-290mA 6.2V
Internal battery status. This shows the approximate battery charge as a percentage, a charge (↑) or
discharge (↓) indicator, battery current (negative=discharging) and the battery terminal voltage. NC
is displayed if the internal battery is not connected.
This screen is not present on models with no internal battery charger, e.g. DT82E
Telstra 3■□□□
203.112.100.109
Internal modem status. While the modem is connected the display shows the mobile carrier name,
network type (e.g. '3' for 3G), signal strength and IP address. If the modem is not connected then
modem status is shown e.g. Modem is off.
This screen is only present on models with an integrated modem (DT8xM)
Eth IP: Auto
192.168.11.160
Ethernet port mode (Automatic, Manual or Disabled) and current IP address
Boiler Temp
97 °C
First user channel (user defined channel name)
Channel 2LM35
17.9
°
C
Second user channel (default channel name)
Alarm4 OFF
1356.3 mV
Alarm #4 state
Oven Over Temp
117.2 °C
Alarm #7 state (alarm text replaces channel name when alarm is active)
Note: channel 1CV is not displayed because it is defined as a working (W) channel. Working channels are neither logged, returned nor
displayed.
Bar Graph
The channel value can be shown as a bar graph instead of a numeric value by using the BG channel option. The BG
option allows the values to be set that represent the left and right side of the graph scale. The channel label can be used
to set the graph scale labels. For example:
Display screen
Comments
E--Fuel Level--F
███████
4V("E--Fuel Level--F",BG10:900)
displays zero scale (no bars) if measured voltage < 10mV; displays full scale (16 bars) if voltage >
900mV
Controlling what is shown on the display
All defined channels and alarms will be shown on the display, except for:
channels which specify the
ND (no display) channel option
working channels (
W channel option)
un-numbered alarms
alarms where the channel in the alarm condition is a no-display (
ND) or working (W) channel
For example:
4TT 'displayed
3CV(W)=3CV+1 'not displayed
ALARM1(2CV>10)"Quokka" 'displayed
ALARM(2CV>20)"Platypus" 'not displayed