Instructions
UM-0085-B09 DT80 Range User Manual Page 267
RG
Section Name
Key Name
Legal Values
Factory Default
Comment
COMMAND_SERVER
SESSION_TIMEOUT
0-65535
0
If 0 the session never disconnects.
Any value can be set in seconds and
command server session will end after
inactivity for the specific period.
HTTP_SERVER PORT
0-65535
80
TCP port number used by web server
(0=disable)
HTTP_SERVER ENABLE_WDG YES,NO
YES
If YES, HTTP and socket monitor is
enabled.(NO=disable)
MODBUS_SERVER
TCPIP_PORT
0-65535
502
TCP port number used by Modbus
server (0=disable)
SERSEN_ADDRESS 0-247
0
The DT80's slave address when using
Modbus on the serial sensor port
(0=disable)
HOST_ADDRESS 0-247
0
The DT80's slave address when using
Modbus on the host port. (0=disable)
USB_ADDRESS 0-247
0
The DT80's slave address when using
Modbus on the USB port. (0=disable)
NTP
BACKGROUND_ENABLE
YES, NO
NO
Periodically synchronise DT80 system
time to an NTP server
SERVER
string
0.datataker.
pool.ntp.org
NTP server address (numeric IP
address or host name)
BACKGROUND_PERIOD
time
30M
-
7D
3599S
Interval between NTP updates
MAX_SLEW_CORRECTION
time 0-60S
3S
Max allowable time adjustment using
"gradual" adjustment method
MAX_JUMP_CORRECTION
time 0-24H
24H
Max allowable time adjustment using
"time jump" method
MIN_CORRECTION
time
0
-
24H
50T
Minimum time adjustment to apply
SLEW_RATE
1-50
10
Percentage change to clock rate during
gradual adjustment
TIMEOUT
time 0-10S
2S
NTP request timeout
BACKGROUND_WAKEUP_DE
LAY
time 0-10M
3S
Delay following wakeup before first NTP
request
LOCALE
TIME_ZONE
time
-24H
-
24H
0S
Local time zone offset
STARTUP
RUN
CURRENT_JOB
,
NONE
, jobname
CURRENT_JOB
Job to be automatically loaded following
hard reset
MAINTAIN_OUTPUTS
YES, NO
NO
Restore state of digital outputs 1..8DSO
and 1RELAY following hard reset
UNLOAD FTP_RETRIES
0-1000, INFINITE
INFINITE
Number of times to retry a failed FTP
unload.
DISPLAY
AUTOSCROLL_INTERVAL
time 0-24H
0S
Auto scroll through channel screens on
LCD at this rate (0=disable)
AUTOSCROLL_DELAY
time 0-24H
30S
Suspend autoscroll for this time if key
pressed
AUTOACK_DELAY
time 0-24H
0S
Auto-acknowledge pop-up messages
on LCD after this time (0=disable)
FUNCTION
Fn_LABEL
string
Label for LCD function key #n (n=1-10)
Fn_COMMAND
string
Command string for LCD function key
#n (n=1-10)
SWITCHES
A, B, C,…Z OFF, ON
varies with switch
Similarly for switches
See Table 8: DT80 Switches (263)
PARAMETERS
Pn
integer
limits vary with
parameter
varies with parameter
Sets the specified parameter to the
specified value, and makes it the
"default" value, i.e. the value to which it
will be set on soft or hard reset.
See Table 7: DT80 Parameters (262).
SMTP_SERVER PORT
0-65535
25
TCP port number used by web server
(0=disable)
WIFI
ENABLE
YES,NO
NO
Enable WIFI Interface on a logger
OPER_MODE
ACCESS_POINT,
CLIENT
ACCESS_POINT
Selects a logger operation mode
defining its role in the WIFI Network