Instructions

UM-0085-B09 DT80 Range User Manual Page 385
RG
CEM20 Specifications
Interfaces
Power/Control inputs: 12Vdc, ground, serial data, serial clock (from logger or previous CEM20)
Power/Control outputs: 12Vdc, ground, serial data, serial clock (to next CEM20)
Analog output: (to logger channel corresponding to CEM20 address)
20 x 4-wire analog inputs
Connections
Max total cable length between data logger and furthest CEM20: 100m
Max supported CEM20 units per data logger:
DT80 (Series 2 or later): 5
DT85 (Series 2 or later): 16
DT81/ 82: none
Sampling
Type: Relay multiplexer
Maximum Input Voltage: 30Vdc
Maximum Sampling Speed: 15Hz
Internal RTD reference junction sensor for thermocouple measurements
System
Status LED: Sample activity
Address Selection: 4-way DIP switch. Address 1-15
Power Supply
Recommended: Logger’s switched 12V output
Alternative: External regulated 12Vdc ± 5%
Power Consumption
Sampling: 0.36W (12V 30mA)
Idle: 0.01W (12V 1mA)
Note: In most cases idle power consumption will be zero because the DT80's 12V output will be turned off when not sampling
Physical and Environment
Construction: Powder coated steel and anodized aluminium
Dimensions: 180 x 100 x 50mm
Weight: 0.55kg
Temperature Range: -40°C to 70°C
Humidity: 85% RH, non-condensing
Accessories Included
Analog and power/control cables for connection to the DT80