RG
Error Number – Description
• Cause/Action
Error Category
Syntax Operation Memory Reading Hardware
E37 – No current job
• A command was entered that operates on the current
job, but there is no job currently loaded.
E39 – Channel list fixed /F
• Changes to program are not allowed
E42 – USB device not ready
• No USB memory device inserted
• DT80 has not yet read the required system information
from the device (wait a few seconds)
• USB device is faulty or not a memory device
E49 – Job has logged data/alarms
• A job that has logged data cannot be deleted. Delete
the data first.
E50 – Job locked
• A job that has been locked cannot be modified
E51 – ALARM/IF command error
• Setpoint character <, >, <> or >< missing
• AND, OR, XOR incorrectly entered
• Setpoint not specified or too large
• Delay incorrectly specified
E52 – Text memory full
• Memory for storage of alarm and expression text is full
E54 – Expression error
• Syntax error
• Expression too complex
E65 – ALARM undefined
• Alarm n does not exist
E74 – Serial sensor CTS detect timeout
• Serial sensor: CTS did not go to the required state
within timeout period
E75 – Serial sensor transmit timeout
• Serial sensor was not able to transmit (e.g. due to flow
control state) within the timeout period
E80 – Serial device not responding
• no response received from SDI-12 sensor
• check cabling and sensor address
E81 – Serial device invalid response
• garbled response received from SDI-12 sensor
• check for address conflict
• check for electrical noise
E82 – Serial device data not available
• SDI-12 sensor doesn't support the requested
measurement
• sensor doesn't support continuous mode
E89 – Serial sensor receive timeout
• Serial sensor: expected characters were not received
within timeout period
E90 – Serial sensor scan error
• Serial sensor: could not convert the received text as
specified in the control string
E101 – Undefined reference: name
• No channel was found with the specified name
E102 – Storefile record size too large
• There are too many logged channels in the schedule
E103 – Invalid storefile: filename
• The indicated storefile appears to be corrupted
E104 – Drive format failed
• Unable to format the specified drive. It may be write-
protected or damaged.