Instructions

UM-0085-B09 DT80 Range User Manual Page 400
RG
Error Number – Description
Cause/Action
Error Category
Syntax Operation Memory Reading Hardware
E127 - Modbus IP address specified on serial
channel
E128 - Modbus serial address specified on
TCP/IP channel
or no address specified
E129 Modbus - unexpected format of response
packet
E130 - Modbus - unexpected function id in
response
E131 - Modbus - exception response received
Slave device does not support the requests used by the
DT80.
E132 - Modbus - error in definition of CV
block to send
invalid CV indEX
E133 - Modbus - reading values to CVs
prevents writing
MODBUS channel cannot write values from CVs and
also assign to CVs
E134 - Modbus - 32 bit format not applicable
to discrete transfer
MBL, MBF options are not applicable when reading 1-
bit registers (coils/discrete inputs)
E135 - Modbus - volume of data to transfer
exceeds Modbus message capacity
limit is 123 16-bit registers or 1968 1=bit registers
E137 Unresolved host name: name
host name may be misspelled
check that DNS server is functional
E142 Modem command failed
MODEM command failed, possibly because the
required prerequisites were not satisfied e.g. PIN not
entered
E143 Modem busy
MODEM command failed due to the modem being
busy, e.g. firmware upgrade in progress
E148-152 Time trigger detailed msg
Time of day schedule trigger syntax error
E153 Create user detailed msg
User account cannot be created as the username
already exists
E154 Create user detailed msg
User account cannot be created as the username
already exists
E153 No such user detailed msg
User account does not exist
E155 Delete user detailed msg
Cannot delete logged in user account
E156 User not authorised detailed msg
Not authorised to perform action
E157 User not authenticateddetailed msg
User cannot be authenticated
E158 Another user logged in
detailed msg
Another user session is active.
Only one session in time is supported
E159 Invalid username detailed msg
The username is invalid